In the case of an alarm there is an optical display via
alarm-LED or an acoustic message via buzzer (if active).
At the same time the alarm is taken over in the
diagnosis and can be read-out after the program end in
the point SETUP / DIAG.
A message to outside can be sent over a contact without
potential via the SUB-D socket (see point 3.4.3).
In pressing the [ENTER] key the optical and acoustic
messages are acknowledged as well as the contact
without potential is set back.
In opening the cap the shaking drive is stopped and the
heating is switched off, independent of the fact, if the
door contact alarm is ON or OFF. If the cap is closed
again, the shaking drive and the heating are operating
automatically again.
In the case of voltage loss during the program
processing or in the operation mode „RUN" the time
which has already been processed is safed. If there is
voltage again, the device starts automatically. The
remaining time of the program is processed. This
happens independent of the fact if the voltage loss
alarm is active or not.
The alarm definition for the voltage loss serves only for
the activation of the optical/acoustic alarm message or
the entry in the diagnosis after the voltage has returned.