Essential Performance
The monitor has the following essential performance in an environment of electromagnetic environment
specified below:
Operating mode
Accuracy
Function
Alarm
Warning
Use of the monitor adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment and
other equipment should be observed to verify that they are operating normally.
Use of accessories, transducers and cables other than those specified or provided by the
manufacturer of the monitor could result in increased electromagnetic emissions or
decreased electromagnetic immunity of the monitor and result in improper operation.
Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of the
monitor, including cables specified by the manufacturer. Otherwise, degradation of the
performance of the monitor could result.
In operation mode, the monitor can be used with electro-surgical unit. The monitor's
operator should ensure the safety of the patients if in use with electro-surgical unit in
accordance with the instructions of this manual. After the elimination of high frequency
signal and high frequency electromagnetic field, the monitor waveform and parameter can
recover within 10 seconds without losing any stored data.
Do not use the monitor with electro-surgical unit in non-operation mode, nor with
large-scale electrical equipment such as ultrasonic, radiation and magnetic resonance
imaging, which may cause electromagnetic interference to the monitor or harm the
monitor's operator.
B EMC Compliance
163