Lid With Screw Cap; Centrifugation; Centrifugation In Continuous Operation - Hettich Universal 320 R Mode D'emploi

Masquer les pouces Voir aussi pour UNIVERSAL 320 R:
Table des Matières
Les langues disponibles

Les langues disponibles

Opening
6.6.4

Lid with screw cap

Fig. 21: Biosafety system
1
Lid
2
Bucket
Closing
Opening
6.7

Centrifugation

6.7.1

Centrifugation in continuous operation

AB1401en
4.
For transport or when inserting and removing the bucket, swing the
bucket into the carrying position (
bracket.
�� The tightness of the biosafety system is also guaranteed in the
carrying position.
Do not rock the biosafety system back and forth during transport,
otherwise leak-tightness is no longer ensured.
1.
Swivel the bracket to the "open" position (
The arrows of the labelling must point downwards so that the word
"open" is legible.
2.
Remove the lid from the bucket.
1
2
1.
Place the lid (
2.
Turn the lid (
1.
Turn the lid (
2.
Remove the lid (
Personnel:
Trained user
1.
Set minutes and seconds at
program.
[START/IMPULS] button.
2.
Press the
�� The centrifugation run is started.
[START/IMPULSE] button lights up during the centrifugation
The
run.
The timing starts at
The rotor speed or the RCF value, the temperature in the centri-
fuging chamber (only for centrifuges with cooling) and the elapsed
time are displayed during the centrifugation run.
[STOP/OPEN] button to cancel the centrifugation run.
3.
Press the
Ramp-down takes place with the set brake level. The brake level is
displayed.
An audible signal sounds when the rotor comes to a standstill.
'OPEN' 'OEFFNEN' is displayed.
Rev.: 13 / 11.2023
4  ) and hold the bucket by the
1  ) centrally on the bucket ( 2  ).
1  ) clockwise until it is tightly closed.
1  ) anticlockwise until it is open.
1  ) from the bucket ( 2  ).
'¥' or retrieve a continuously running
'00:00' .
Operation
1  ).
29 / 44