6.2 CLEANING the Main Unit
To clean the Main Unit(1) use a cloth slightly moistened
with an antibacterial detergent, then dry it with a soft
paper towel.
WARNING
Do not immerse the Main Unit (1) in water and
do not wash it under running water.
6.3. SANITISATION
6.3.1. Immerse for at least 30 minutes the
opened Removable Medication Chamber
(2) and the accessories 2G-3-4-5, in a solu-
tion of 50% water and 50% white vinegar,
then rinse thoroughly with hot running wa-
ter.
6.1.2. DRYING
6.1.2.a. After washing and rinsing all the com-
ponents, shake them to remove excess wa-
ter and reassemble them, as described in
the following paragraph 7. REASSEMBLY
OF THE DEVICE. At this point, to eliminate
the residual water that has deposited in the
nebuliser head (2G), press Button 1B (side
figure) and wait for the device stop to nebu-
lizing. Turn off the unit. This operation is very
important, because it prevents the deposit
of limestone in the micropores of the Nebu-
liser Head (2G) which could affect proper
nebulisation of the drug in subsequent ther-
apeutic applications.
2
2
6.1.2.b. To complete the drying of all the com-
ponents, leave them out and in a dry place
(e.g. not in the bathroom) as into the side
figure.
2G
2G
12