Setting Filter; Selecting Bmi/Bsa/Waist Circumf - Seca 651 Mode D'emploi

Masquer les pouces Voir aussi pour 651:
Table des Matières
Les langues disponibles

Les langues disponibles

Selecting BMI/BSA/Waist circumf.

Setting filter

The Filter function can be used to avoid interferences during weight determi-
nation. The selected setting has the following influences on measurements
with the Hold/Autohold function activated:
• Sensitivity with which the weight display reacts to patient movements
• Period of time until a weight value is displayed permanently.
1. Press the
The Settings menu is displayed.
2. Press the
3. Press the Filter item.
The current setting is displayed.
4. Press the desired filter level.
The setting is active.
You can specify whether the device calculates the Body Mass Index (BMI) or
the Body Surface Area (BSA) as soon as the patient's weight and height are
available.
The device can also be set so that the waist circumference (Waist circumf.)
can be entered manually.
Device mode
Basic
Advanced
Expert
Service
key.
or
key until the Filter menu item is displayed.
Filter
Low
Medium
High
NOTE
With the Low setting, it may occur that no weight value is displayed
permanently despite the activated Hold function in the case of
patients who are not very steady on their feet.
Device mode
Basic
Advanced
Expert
Service
Available
Weight
determination
Fast
Medium
Slow
Function available
Configuration • 91