Check the 4 ... 20 mA signal
Connect a multimeter in the suitable measuring
range according to the wiring plan.
Error
Cause
4 ... 20 mA
No atmospheric
signal not
pressure compen-
stable
sation
4 ... 20 mA
Wrong connection
signal mis-
to voltage supply
sing
No power supply
Operating voltage
too low, load resis-
tance too high
In Ex applications, the regulations for the
wiring of intrinsically safe circuits must
be observed.
5.3
Shorten suspension cable
Shorten the suspension cable individually.
Proceed as follows:
1. Remove the filter adapter from the capillary
line
2. Cut the suspension cable to the requested
length with side cutters
Caution:
Do not squeeze the capillary cable, as
this will impair the pressure compensati-
on. If necessary, rework the capillary with
a sharp knife.
3. Remove approx. 10 cm of the cable mantle,
strip off approx. 1 cm of insulation from the
ends of the wires
4. Insert the filter adapter
The work steps are finished.
VEGAWELL S 51 •
5.4
You can find an instrument return form as well
as detailed information about the procedure in
Rectification
the download area of our homepage. By doing
Check the capilla-
this you help us carry out the repair quickly and
riy and cut it clean
without having to call back for needed informa-
tion.
Check the pressu-
In case of repair, proceed as follows:
re compensation
and clean the filter
element, if neces-
sary
Check connec-
tion according to
chapter "Connec-
tion steps" and if
necessary, cor-
rect according to
chapter "Wiring
plan"
6
Check cables for
breaks; repair if
necessary
6.1
Check, adapt if
necessary
Take note of chapters "Mounting" and "Connec-
ting to voltage supply" and carry out the listed
steps in reverse order.
6.2
Pass the instrument on to a specialised recy-
cling company and do not use the municipal
collecting points.
Remove any batteries in advance, if they can be
removed from the device, and dispose of them
separately.
If personal data is stored on the old device to be
disposed of, delete it before disposal.
If you have no way to dispose of the old instru-
ment properly, please contact us concerning
return and disposal.
How to proceed if a repair is
necessary
Print and fill out one form per instrument
Clean the instrument and pack it damage-
proof
Attach the completed form and, if need
be, also a safety data sheet outside on the
packaging
Ask the agency serving you to get the
address for the return shipment. You can find
the agency on our homepage.
Dismount
Dismounting steps
Warning:
Before dismounting, be aware of dan-
gerous process conditions such as e.g.
pressure in the vessel or pipeline, high
temperatures, corrosive or toxic media
etc.