Technical data

Transmitter
Description
Enclosure
Ambient temperature
Storage temperature
Installation
Mechanical vibration
Design
Power supply
Battery change interval
Display
Push button
Measuring function
Communication
Digital output
Pulse output A
Pulse output B
Pulse width
Max. pulse frequency
Volume units
Flow units
Alarm codes
Cable length
EMC
Weight
Sensor
MID: minimum temperature 15°C (59°F)
1)
Beschreibung
Pipe design
Nominal size
Pressure rate
Pipe material
Transducer design
Transducer material
Media temperature
MID: environment class -10°C...+55°C (14°F....131°F)
2
*
Waste industrial batteries are accepted back by the producer or importer, who has originally marketed the battery, or by the produ-
cer or importer, where the new industrial battery is purchased.
34
Specification
IP67 according to EN 60529 and DIN 40050 (NEMA 4X/6)
0 °C ...60 °C (32 °F.... 140 °F)
-35 °C...85 °C (-40 °F...185 °F)
Cable max. 5, 10, 20, 30 m (16.4, 33, 65, 90 ft) from sensor
2 g, 1...800 Hz sinusoidal in all directions to IEC 68-2-6
Fibre glass reinforced polyamide in light-gray color
• Battery: replaceable 3.6 V LiSOCl (Lithium Thionyl Chloride)
battery pack 32 Ah
• Mains: 87 ... 265 V AC (50 ... 60 Hz)
6 years at 60 °C (140 °F) operation
LCD, 8 digits, additional 2 digits and symbols for status information
One push button for toggling between display information
0.5 Hz battery mode or 20 Hz mains-powered
IrDA on display panel (MODBUS RTU protocol);
separate add-on serial interface moduls RS232 or RS485
(also MODBUS RTU protocol)
Two passive, galvanically isolated open drain-mos outputs A and B
Max. ±35 V, 50 mA
Preset to pulse output for forward flow
Preset to alarm for present failure
5, 10, 20, 50, 100, 200, 500 ms
100 Hz
m
3
m
/h (default)
3
Track 1, 2 measuring, internal failure, battery low, flow overload,
pulse output frequency overload
Max. 30 meter between transmitter and pipe (factory sets: 5, 10, 20, 30 m)
Emission EN 61000-6-4
Immunity EN 61000-6-2
Transmitter:
1.5 kg (3 lb)
Technische Daten
2-track sensor with flanges and integrated transducers wet-calibrated from factory
DN 50, 65, 80, 100, 125, 150, 200, 250, 300, 350, 400, 500, 600, 700, 800, 900, 1000,
1200
PN 16, PN 25, PN 40, EN 1092-1
DN 50 ... 80:
Bronze
DN 100 ... 1200: Carbon Steel EN 1.0345 / p235 GH, painted in light-gray
DN 50 ... 80:
Mounted in sensor.
DN 100 ... 1200: Integrated version welded onto pipe.
Stainless steel (AISI 316 / 1.4404) / brass (CuZn36Pb2As)
Compact:
DN 50 ... 1200: 2 ... 120 °C (35.6 ... 248 °F)
Remote:
DN 50 ... 80: 2 ... 150 °C (35.6 ... 302 °F)
DN 100 ... 1200: 2 ... 200 °C (35.6 ... 392 °F)
2)
1)
1)
1)
AMFLO