Télécharger Imprimer la page

Agilent Technologies X3500-64000 Mode D'emploi page 234

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 43
16
Instructions for Use
Pump Fixing
Pump Fixing
Failure to comply with these installation instructions could result in the pump
WARNING!
detaching from the system in the event of a rotor failure, which could cause
property damage or serious injury or death.
The TwisTorr 304 FS can be installed in any position. Fix the
TwisTorr 304 FS in a stable position, mounting the inlet flange of the
turbopump to the system counter-flange, with a connection capable
of withstanding a torque of 1000 Nm around its axis.
The ISO-K flange can be fixed using high strength steel clamps (like
Agilent model IC63250DCMZ).
The following table shows, for each flange, the necessary number of
IC63250DCMZ clamps and the relevant fixing torque.
Tab. 2
FLANGE
ISO 100 K
ISO 160 K
The turbopump with ConFlat inlet flange must be fixed to the
vacuum chamber by means of the appropriate Agilent hardware. See
the appendix "Technical Information" for a detailed description.
The TwisTorr 304 FS cannot be fixed by means of its base.
NOTE
234/304
FIXING DEVICE
M10 clamps
M10 clamps
TwisTorr 304 FS User Manual / 87-901-027-01(B)
N.
FIXING TORQUE
4
22 Nm
4
22 Nm

Publicité

loading