Instructions for Use
Safety Guideline for Turbomolecular Pumps
Safety Guideline for Turbomolecular Pumps
Turbomolecular pumps as described in the following operating
manual contain a large amount of kinetic energy due to the high
rotational speed in combination with the specific mass of their
rotors.
In case of a malfunction of the system for example rotor/stator
contact or even a rotor crash the rotational energy may be released.
To avoid damage to equipment and to prevent injuries to operating personnel
WARNING!
the installation instructions as given in this manual should be strictly
followed!