Bibby Scientific Stuart SA6 Manuel D'utilisation page 3

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Safety Advice Before Use
For your own safety and the safety of the
mixer please read the following safety
instructions before using the Mini Vortex
Mixer.
v Gently take the Mini Vortex Mixer out of the
packaging and place it on a flat, stable,
clean, non-slippery surface.
v Do not lift or hold the device by the cable.
v Never move or carry the unit when in use or
connected to the mains electricity supply.
v The Mini Vortex Mixer should be used for the
specified application only. It must not be
operated in a hazardous or flammable
environment and must not be used to mix
explosive or highly reactive material.
v Overfilling of liquid may cause the medium
to spill out. You should therefore always use
covered test tubes.
v CAUTION! The equipment will restart, both
in the case of mains interruption and in the
case of any fault or mechanical interruption.
Care should be taken based on the sample
used for mixing.
v For best results only use to mix tubes up to
30mm diameter.
Operation
Position the apparatus on a firm level surface.
Operation starts automatically when the soft
rubber cup is pressed down by the test tube, and
stops once the test tube is removed. Before use
ensure that the speed control knob is turned
fully anti-clockwise. Rotate speed knob to
increase or decrease the speed between
2500-4500rpm. Rotate clockwise to increase the
speed and anti-clockwise to decrease the speed.
Maintenance, Cleaning
& Servicing
WARNING: Ensure the unit is
disconnected from the mains electricity
supply before attempting maintenance
or servicing.
The whole unit should be cleaned regularly with
a soft, damp cloth and mild detergent solution.
Do not use harsh or abrasive cleaning agents.
Excessive amounts of liquid should be avoided.
Liquid should not come in contact with the
motor.
After cleaning, ensure that all parts are dried.
Do not use the unit if it shows signs of
mechanical or electrical damage. Do not use the
equipment if the rubber cup or suction cup are
damaged.
The brushless motor in the Mini Vortex Mixer
does not require any routine maintenance.
Any required servicing should be performed
by authorised, qualified personnel only.
The following spares are available from your
laboratory dealer:
Description
Spare Rubber Cup
For a comprehensive list of parts required by
service engineers conducting internal repairs and
a service manual, please contact the Technical
Service Department of Bibby Scientific Ltd.
quoting both the model and serial number.
Only spare parts supplied or specified by Bibby
Scientific Ltd. or its agents should be used. Fitting
of non-approved parts may affect the
performance and safety features designed into
the instrument.
If in any doubt, please contact the Technical
Service Department of Bibby Scientific Ltd. or the
point of sale:
Bibby Scientific Ltd.
Stone, Staffordshire, ST15 0SA
United Kingdom
Tel: +44 (0)844 936 0234
Fax: +44 (0)844 936 0235
E-mail: info@bibby-scientific.com
www.bibby-scientific.com
Catalogue Number
SA6/1
3

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Stuart mini vortex mixer

Table des Matières