Shut-Down Procedures (end of day)
1. Flush the
Dentsply
system according the dental unit manufacturer's
instructions.
2. Remove any ultrasonic inserts and the Steri-Mate
handpiece. Clean and sterilize the handpiece and
all inserts used during the procedure (refer to
the Infection Control information booklet and the
Cavitron Ultrasonic Inserts Directions for Use for
cleaning and sterilizing instructions).
3. Clean and disinfect the handpiece cable assembly
by applying a medically approved non-immersion
type disinfectant solution,* carefully following the
instructions provided by the solution manufacturer.
To clean the cable, generously spray disinfectant
solution on a clean towel and wipe the cable and
connector. Discard used towel. To disinfect the
system, generously spray disinfectant on a clean
towel and wipe the cable and connector. Allow the
disinfectant solution to air dry.
4. Clean and disinfect all surfaces of the dental control
head according to the manufacturer's instructions.
5. Place the cable connector into the handpiece holder
for storage.
6. Turn off the power and water to the dental unit.
*NOTE: Water-based disinfection solutions are preferred.
Some alcohol-based disinfectant solutions may discolor the
handpiece cable and connector.
ENGLISH • 12
88016 Rev (0423)_ML.indd 12
88016 Rev (0423)_ML.indd 12
Cavitron built-in ultrasonic
Handpiece Cable
While the Cavitron Handpiece Cables are designed
to offer optimum flexibility and endurance, improper
handling may lead to a shortened useful life of the
cable assembly, due to excessive wear on the wires
and/or internal tubing. To prevent premature failure of
the cable assembly, please use the following steps:
1. Do not wrap the cable tightly around any apparatus.
2. Avoid overstretching or pulling the cable.
3. If possible, minimize frequent bending/folding that
exceeds normal usage during dental procedure.
4. For detachable handpiece cables, hold by the plug
body when detaching from the Cavitron system.