10.3 Patient Comfort Considerations
Reasons for sensitivity
• Incorrect tip placement. Point should be directed
away from root surfaces.
• Allowing the insert to remain in a static position on
the tooth. Keep the tip in motion and change it's
path on the tooth.
• Applying pressure. Use extremely light grasp and
pressure, especially on exposed cementum.
• If sensitivity persists, decrease power setting and/
or move from the sensitive tooth to another and
then return.
Section 11: System Care
Start-up Procedures (beginning of the day)
1. Switch on the dental unit (refer to dental unit
instructions for use).
2. Remove the Dentsply Sirona Cavitron cable
connector assembly from the handpiece holder and
set the power adjustment to 30% of maximum.
3. Connect Steri-Mate handpiece to the cable
connector assembly (refer to the Infection Control
information booklet for sterilizing instructions).
4. Adjust the lavage control to maximum.
5. Hold the handpiece (without an insert installed)
over a sink or drain. Activate the foot control and
flush water through the handpiece for at least two
minutes.
6. Place a sterilized insert into the handpiece using a
gentle push-twist motion.
88016 Rev (0423)_ML.indd 11
88016 Rev (0423)_ML.indd 11
7. Activate the ultrasonics and adjust the power and
lavage control to your preferred operating positions.
Between patients
1. Remove any ultrasonic inserts and the Steri-Mate
handpiece. Clean and sterilize the handpiece and
all inserts used during the procedure (refer to
the Infection Control information booklet and the
Cavitron Ultrasonic Inserts Directions for Use for
cleaning and sterilizing instructions).
2. Clean and disinfect the handpiece cable assembly
by applying a medically approved non-immersion
type disinfectant solution,* carefully following the
instructions provided by the solution manufacturer.
To clean the cable, generously spray disinfectant
solution on a clean towel and wipe the cable and
connector. Discard used towel. To disinfect the
cable, generously spray disinfectant on a clean
towel and wipe the cable and connector. Allow the
disinfectant solution to air dry.
3. Clean and disinfect all surfaces of the dental
treatment control head according to the
manufacturer's instructions.
4. Connect a sterilized Steri-Mate handpiece onto its
corresponding cable connector.
5. Hold the handpiece (without an insert installed)
over a sink or drain. Activate the foot control and
flush water through handpiece for at least thirty
seconds.
6. Place a sterilized insert into the handpiece.
ENGLISH • 11