Technical specifications
Units are expressed in cm H
90W power supply unit
AC input range:
DC output:
Typical power consumption:
Peak power consumption:
Environmental conditions
Operating temperature:
Operating humidity:
Operating altitude:
Storage and transport temperature:
Storage and transport humidity:
Electromagnetic compatibility
The AirSense 10 complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC60601-1-
2:2007, for residential, commercial and light industry environments. It is recommended that mobile communication devices
are kept at least 1 m away from the device.
Information regarding the electromagnetic emissions and immunity of this ResMed device can be found on
www.resmed.com, on the Products page under Service and Support.
IEC 60601-1:2005 classification
Class II (double insulation), Type BF, Ingress protection IP22.
Sensors
Pressure sensor:
Flow sensor:
Maximum single fault steady pressure
Device will shut down in the presence of a single fault if the steady state pressure exceeds:
30 cm H
O (30 hPa) for more than 6 sec or 40 cm H
2
Sound
Pressure level measured according to EN ISO 17510-1:2009 (CPAP mode):
SlimLine:
Standard:
Power level measured according to EN ISO 17510-1:2009 (CPAP mode):
SlimLine:
Standard:
Declared dual-number noise emission values in accordance with ISO 4871:1996.
16
All manuals and user guides at all-guides.com
O and hPa. 1 cm H
O is equal to 0.98 hPa.
2
2
100–240V, 50–60Hz 1.0–1.5A, Class II
115V, 400Hz 1.5A, Class II (nominal for aircraft use)
24V
53W (57VA)
104W (108VA)
+5°C to +35°C
Note: The air flow for breathing produced by this therapy
device can be higher than the temperature of the room.
Under extreme ambient temperature conditions (40°C) the
device remains safe.
10 to 95% relative humidity, non-condensing
Sea level to 2,591 m; air pressure range 1013 hPa to
738 hPa
-20°C to +60°C
5 to 95% relative humidity, non-condensing
Internally located at device outlet, analogue gauge pressure
type, -5 to +45 cm H
Internally located at device inlet, digital mass flow type, -70
to +180 L/min
O (40 hPa) for more than 1 sec.
2
26.6 dBA with uncertainty of 2 dBA
26.6 dBA with uncertainty of 2 dBA
34.6 dBA with uncertainty of 2 dBA
34.6 dBA with uncertainty of 2 dBA
3.75A
O (-5 to +45 hPa)
2