Error
Description
Number
–180
"Macro error"
–181
"Invalid outside macro
definition"
–183
"Invalid inside macro
definition"
–184
"Macro parameter
error"
–200
"Execution error"
–201
"Invalid while in local" Indicates that a command is not executable while
–202
"Settings lost due to rtl" Indicates that a setting associated with a hard local
–203
"Command protected" Indicates that a legal password-protected program
FTB-1v2 and FTB-1v2 Pro
Probable Cause
This error, as well as errors –181 through –189, are
generated when defining a macro or executing a
macro. This particular error message should be
used if the device cannot detect a more specific
error.
Indicates that a macro parameter placeholder
($<number) was encountered outside of a macro
definition.
Indicates that the program message unit sequence,
sent with a *DDT or *DMC command, is
syntactically invalid (see IEEE 488.2, 10.7.6.3).
Indicates that a command inside the macro
definition had the wrong number or type of
parameters.
This is the generic syntax error for devices that
cannot detect more specific errors. This code
indicates only that an Execution Error as defined in
IEEE 488.2, 11.5.1.1.5 has occurred.
the device is in local due to a hard local control
(see IEEE 488.2, 5.6.1.5); for example, a device with
a rotary switch receives a message which would
change the switches state, but the device is in local
so the message can not be executed.
control (see IEEE 488.2, 5.6.1.5) was lost when the
device changed to LOCS from REMS or to LWLS
from RWLS.
command or query could not be executed
because the command was disabled.
SCPI-Based Errors
447