Using Your Unit and Modules in an Automated Test Environment
Standard Status Data Structure
The following table presents a summary of the possible operations on STB
and SRE registers.
Register
STB
Use *STB?.
Use serial poll (GPIB
bus sequence that
allows retrieval of the
value without
interrupting the current
process).
SRE
Use *SRE?.
The diagram displayed on the next page is a useful aid in understanding
the general commands and how a service request (SRQ) is generated.
Using a service request, a device notifies the controller that an event
requiring special attention occurred. The controller will then find which
device generated a SRQ (its RQS bit is set) and the causes of it.
264
Read
Impossible to write;
the register content is
only modified when
the Event registers or
Queues are modified.
Use *SRE with a value
equal to 0 to disable
the register or with a
value equal to 1 to
enable it.
Write
Use *CLS before
sending a query (to
clear the Event registers
and Queues and by the
same token clear the
STB register).
Clear
Use *SRE with a
value equal to 0.
At startup, the
register is set to 0.