Télécharger Imprimer la page

arjo Alpha Active 3 Manuel D'utilisation page 29

Masquer les pouces Voir aussi pour Alpha Active 3:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Guidance and manufacturer's declaration - electromagnetic immunity
Voltage dips, short
0 % UT; 0,5 cycle
interruptions and
At 0°, 45°, 90°,
voltage variations
135°, 180°, 225°,
on power supply
270° and 315°
input lines
0 % UT; 1 cycle
IEC 61000-4-11
and
70 % UT; 25/30
cycles
Single phase: at 0°
0 % UT; 250/300
cycle
NOTE
UT is the AC mains voltage prior to application of the test level.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
a)
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the product is used exceeds the applicable RF compliance level above, the product
should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 1 V/m.
b)
0 % UT; 0,5 cycle
Mains power quality should be that of
At 0°, 45°, 90°,
a typical commercial or hospital
135°, 180°, 225°,
environment. If the user of the pump
270° and 315°
requires continued operation during
power mains interruptions, it is
0 % UT; 1 cycle
recommended that the pump be
and
powered from an uninterruptible
70 % UT; 25/30
power supply or a battery.
cycles
Single phase: at 0°
0 % UT; 250/300
cycle
23

Publicité

loading