To Disconnect the
Tubeset
Transport Mode
Transport mode is non-therapeutic offering support only for up to 12 hours. It
is recommended that when in transport mode the patient is frequently
monitored.
To Deflate and Store the Alpha Active 3 Mattress
To deflate the mattress
To store the mattress
To disconnect the tube-set at any time, depress the buttons on
the top and bottom of the tubeset connector pull the tube-set
connector away from the pump.
To deflate the mattress Refer to Page 11 "To Deflate and Store
the Alpha Active 3 Mattress".
To transport a patient using the Alpha Active 3 mattress overlay,
disconnect the tube-set from the pump and place the attached
cap over the end of the tubeset to put the mattress into transport
mode. This will automatically switch the mattress into transport
mode.
The patient will remain supported by the mattress for up to 12
hours.
To resume normal operation, simply reconnect the tube-set and
run the pump.
Caution
1.
Disconnect the tube-set from the pump.
2.
Activate the CPR control to deflate the mattress.
Following deflation:
1.
Bring the tubeset over the mattress to lie parallel to the foot
end of the mattress.
2.
Roll the mattress from the foot end toward the CPR
connector at the head end of the mattress.