Appendix - Maintenance schedule
Maintenance Schedule Interrliner (Example)
Sample volume: 350 per day (5 working
days)
or earlier after number of samples
Perform End-of-day wash
Clean outside aspiration needle
Check tubing/diluent syringe
Clean outside instrument
Clean Fill nozzle
Clean Liquid separator
Check sensors
Replace Fill nozzle O-ring
Replace bacterial air filter (HEPA)
Replace Disc filter (white) Only if internal
waste container is used
Interrliner
Version 1.01 MRN-164_1-FR
Daily
Weekly
Level 4
(monthly)
8.750
samples
X
X
X
X
X
X
X
X
X
X
Level 3
Level 2
Level 1
(quarterly
(halfyearly
(yearly)
)
)
26.500
52.500
105.000
samples
samples
samples
Annexe pour l'Interrliner
Parts
Total
amount/year
Per 105.000
samples
QWLV050004
12
QWLV040002
12
QWLV040001
12