5. Commissioning, operation
Sensor connection cable
DANGER!
Danger to life due to loss of explosion protection
Non-observance of these instructions and their contents may result in the
loss of explosion protection.
Only carry out set-up work in non-hazardous environments!
Use only the sensor connection cable, approx. 1.1 m (3.3 ft) (Ex version),
see chapter 10 "Accessories". This is marked by a blue plastic ring on the
cable.
Sensor connection cable with blue
plastic ring
Use cable in hazardous area!
WIKA operating instructions, reference pressure sensor, model CPT62I0
Sensor connection cable without blue
plastic ring
Do not use cable in hazardous area!