Maintenance, Cleaning And Recalibration; Maintenance; Cleaning - WIKA CPT62I0 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

6 Faults / 7. Maintenance, cleaning and recalibration
Faults
No output signal
Leakage at the process
connection

7. Maintenance, cleaning and recalibration

Personnel: Skilled personnel
Protective equipment: Protective gloves and safety goggles
Tools: Spanner or torque spanner
For contact details see chapter 1 "General information" or the back page of
the operating instructions.

7.1 Maintenance

This model CPT62I0 reference pressure sensor is maintenance-free.
Repairs must only be carried out by the manufacturer.
Only use original parts (see chapter 10 "Accessories").

7.2 Cleaning

CAUTION!
Physical injuries and damage to property and the environment
Improper cleaning may lead to physical injuries and damage to property and
the environment. Residual media in the dismounted instrument can result in a
risk to persons, the environment and equipment.
Wear the requisite protective equipment.
Carry out the cleaning process as described below.
1. Before cleaning, correctly disconnect the instrument from the pressure supply, switch it
off and disconnect it from the voltage supply (CPH62I0).
2. Use the requisite protective equipment.
WIKA operating instructions, reference pressure sensor, model CPT62I0
Causes
Cable break
Hairline crack in the pressure
port
Measures
Visually check the continuity of
the cable.
If there is any damage to the
cable, replace it with a new
one.
Consult with manufacturer, if
required, replace instrument
EN
15

Publicité

Table des Matières
loading

Table des Matières