EN 61010-1, Protection class I,
Standard power supply: Over voltage protection category II.
Safety
Low voltage power supply: Over voltage category III.
Environment < 2000 m altitude
Installation depth: approx. 210 mm incl. terminals
Panel cut out: 138
Panel thickness: 2...40 mm, fixing to DIN 43834 Weight approx. 3.5 kg,
front bezel/door from die cast metal.
Model/weight
Ingress protection:
front IP 54 (EN 60 529, Cat. 2)
rear IP 20 (EN 60 529, Cat. 2)
field housing IP 62
Keyed plug-on screwed terminal strips.
Connections
Cable cross section area max. 2.5 mm² (with ferrules).
To DIN 19240:
Logic "0" equals -3...+5 V, active with logic "1" (equals +12...+30 V, max. 25 Hz,
4 digital inputs
max. 32 V,
Input current approx. 1.5 mA
Each input with a selectable function: Control input On/Off event.
Auxiliary power
For powering digital inputs with potential free contacts 24 VDC, max. 30 mA,
short circuit protected, unstabilised.
supply
Changeover contact, 230 VAC / 3 A, for alarm condition transmission,
3 relays
at desk top version: max. 30 V
Type (RS 232 / RS 485) and unit address can be selected
Max. cable length using screened cable: 15 m (RS 232) /
Serial interface
1000 m (RS 485)
Galvanically isolated from the system.
Internal Ethernet
Given that this is an interface for industrial use, a screened cable (CAT5) must
interface, 10BaseT,
be used. If a PC is to be connected directly (without Hub or Switch) then a so
plug type RJ45
called cross-over cable must be used.
10BaseT, 10 MBit/s
The following
The devices function according to the 10BaseT standard with 10MBit/s. It is,
standards are
however, also possible to integrate the device into a 100BaseTx network via an
auto-sensing hub or switch. This kind of auto-sensing component sets itself
supported:
automatically to the transmission rates supported by the end device.
Outputs
24 V+/- 20% max. 25 mA (internal current limiter)
Communication
The resistors for HART® communication are built in (250 Ohm); Ø 2 mm sockets
resistance
Galvanic isolation
Test voltage 500 V to all other circuits
+1
+1
x 138
mm.
/ 60 V
eff
DC
8. Technical data
Electrical security
Housing/Installation
Option "Digital I/O"
Option "Ethernet"
Option "Loop power