Please be aware that this number is not arbitrary, rather it must be defined dependent
on the network address of the TCP/IP network. The input form corresponds to syntax
(e.g. 172.016.231.005). Complete the input with "↵ accept".
The subnet mask must be input if the device is to establish connections in another
sub-network. Specify the subnet mask of the sub-network in which the device is
located (e.g. 255.255.255.000). Please observe: the IP address determines the class
of the network. The result of this is a default subnet mask (e.g. 255.255.000.000 for a
Class B network).
Enter the IP address of the gateway here, if connections are to be established in other
networks.
As the unit does not, at the moment, set up a connection itself via the Ethernet, it is
not necessary to enter a Gateway. Leave the set-up at "0.0.0.0".
Note:
Changes to the system parameters are not activated until the SETUP menu is closed
and the settings accepted. Only then will the device work with the new settings.
4. Setting up unit - SET UP
4.5.4 Subnet mask
allocation
4.5.5 Gateway allocation