Specification
Timer:
Fuses:
Safety:
Maximum Continuous Overload:
E.M.C:
Environmental protection:
Temperature range:
Humidity:
Altitude:
Dimensions:
Weight:
Cleaning:
24
0 – 90 minutes: user selects test duration.
FS1: 100 mA (T), 250 V IEC 127/1.
FS2: 100 mA (F), 250 V IEC 127/1.
Mains power cord fused plug (when applicable): 13 Amp fuse to BS1362.
The instrument meets the safety requirements for double insulation to IEC 1010-1 (1995).
EN 61010 (1995) to installation Category III, 300 Volts phase to earth (ground) and 500
Volts phase to phase.
1 kV rms.
The instrument meets EN 50081-1, EN 50082-1 (1992) and EN 61326-1(1998).
IP54 (with the charging recess cover securely closed).
Operating: –20°C to +50°C (Errors double outside range 0˚C to 30°C).
Storage: -25°C to +65°C.
90˚ RH at 40˚C
2000 metres maximum to operate within full specification.
344 mm x 245 mm x 158 mm.
5.6 kg.
Wipe the disconnected instrument with a clean cloth dampened with soapy water or
Isopropyl Alcohol (IPA).