10 PC Diagnostic Tool
10.2 Test procedure
78
Operating Instructions (valid for Canada) CEREC AF, CEREC AF Connect
3. On completion of the test, press any key to go on to the test dialog.
Tip: You can scroll to the individual test steps using the arrow keys.
The corresponding result is shown in front of each test step.
4. To end the test, restart the PC.
Test result: Pass
No errors were found on the PC-specific hardware. Replacing the PC
component or the PC is not advisable.
1. Perform the separate test for the supply board.
2. Perform an image restore with the restore set for troubleshooting.
3. Check the service instructions for other possible fault sources.
Test result: Fail
An error was found on the PC-specific hardware. Replacing the PC
component or the PC may be required.
An image restore is not advisable.
1. Check to find out which components did not pass the test.
2. Carry out the appropriate steps in the chapter entitled
Troubleshooting [ → 80] .
3. Note the number of the failed test and the error code if available.
Attach this information to the returned PC when replacing a PC.
End "Technician Diagnostics" test
1. Remove the PC diagnostic tool CD from the drive.
2. Switch the PC off by briefly pressing the on/off key.