8 Maintenance
8.3 Color calibration
72
Operating Instructions (valid for Canada) CEREC AF, CEREC AF Connect
Ending the color calibration
The software indicates that the color calibration was completed
successfully.
"Ok"
1. Click on the
button.
 The CEREC Omnicam is now color calibrated.
2. Remove the color-calibration set from the camera and place it back
in the packaging.
Error message during color calibration
The software indicates if an error occurs during color calibration. If the
color calibration contained an error, ensure the following:
● The color-calibration set is free of dust
● The color-calibration set was mounted correctly
● The CEREC Omnicam exit window is clean
➢ Then restart the color calibration.
Do not continue using a damaged color-calibration set; instead, contact
your distributor to purchase a new one.
Replacing the color calibration set
Please note that the color calibration set
● can only be used with CEREC SW software ≥ 4.5 or Sirona Connect
SW ≥ 4.5 or CEREC Premium SW ≥ 4.5.
● can only be kept for use for a maximum of 2 years. You can find the
expiry date at the bottom of the color calibration set container.
Previous storage may mean that the period for use has been reduced
to less than 2 years.
● can only be used for one year after the container has been opened.
Write the date that the container was opened on the container after
"Opened on ___________" using a waterproof pen and do not use
after one year.
The color calibration set may no longer be used once either of the two
periods has expired.
The software notifies you that the color calibration set needs to be
replaced with a new set before the color calibration expires.
Once the color calibration set has expired the software notifies you that a
color analysis can only be carried out based on old calibration data.
Please contact your dealer for replacements for the color calibration set.