8 Maintenance
8.1 Care, cleaning, disinfection and sterilization
64
Operating Instructions (valid for Canada) CEREC AF, CEREC AF Connect
13. Carefully and slowly remove the plug with the holder from the mirror
sleeve by downward movement of the holder. Make sure that the
mirror sleeve points upwards not to risk any damage by liquids
entering the inside of the mirror sleeve.
14. Remove the plug from the holder.
15. Dry the plug with a clean lint-free cloth.
16. Prior to each use of the HLD set test the solution for its effectiveness
per manufacturer´s instructions. If not in use rinse container and
holder and store the holder inside the empty container.
17. Dry the whole surface of the mirror sleeve and also the lower part of
the inner surface using a clean soft, lint-free cloth.
18. Store the mirror sleeve in a manner to protect it from contamination
until the time of use.
19. At time of use remove the white protective cap from the camera tip.
20. Carefully re-attach the mirror sleeve and allow it to lock in place (see
„Refitting the mirror sleeve [ → 58]").
Dentsply Sirona
66 37 487 D3652
D3652.201.01.14.02 10.2017