Storing a color-calibration set
The color-calibration set must be stored in its packaging in a dry place
which is protected from light. It must be used with a disinfected Omnicam
as the color-calibration set must itself not be disinfected. If dust
accumulates on the inside of the color-calibration set, it must be carefully
removed using compressed air.
Switch on the color analysis
1. In the software, navigate to the system menu and click on the
"Configuration"
button.
2. Click on the
"Devices"
"Omnicam"
3. Click on the
"Shade Detection"
4. Select the
- You can choose between various color systems (
Selection"
).
- You can decide whether you would like to be notified in 14 days
when the color calibration is needed again.
- If color recognition is not possible with your camera, a
corresponding notice will appear. A color calibration is also not
available in this case.
5. Confirm the changes below with
"Color Calibration"
6. Click the
calibration.
Color-calibrating the camera
1. Make sure that the CEREC Omnicam is clean, disinfected and dry.
2. Remove the color-calibration set from the packaging.
3. Use the CEREC Omnicam to scan the QR code on the underside of
your color-calibration set. In order to do this, you must hold the
CEREC Omnicam still in front of the QR code so that it is completely
visible in the picture. If the QR code appears to be shiny, hold the
camera at more of an oblique angle in order to avoid any glaring light
and to make it easier to scan the codes. If the QR code is recognized,
"Please mount color calibration set"
the next
This step of the QR code scan is skipped during the subsequent color
calibration and the serial number of the color-calibration set is thus
displayed. If this does not match the serial number printed on your
color-calibration set, click on the
the new QR code.
4. Mount the color-calibration set on the tip of the camera until it locks
into place.
"Ok"
5. Click on the
button.
 The measuring process starts. Do not move the CEREC
Omnicam or the color-calibration set during this time.
 The software provides status updates on the calibration and
informs you once the procedure is complete.
button.
button.
option.
"Ok"
.
button and carry out the color
step appears.
"Rescan QR Code"
8 Maintenance
"Shade Guide
button and scan
71