Hach NA5600 sc Na+ Maintenance Et Dépannage page 5

Masquer les pouces Voir aussi pour NA5600 sc Na+:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 29
1. Put on the personal protective equipment identified in the safety data sheet (MSDS/SDS).
2. Push menu, then select STOP ANALYZER.
Note: If START ANALYZER shows, the analyzer is already in standby mode.
3. Replace the reactivation solution and calibration solution as follows:
a. Remove the reactivation solution and calibration solution bottles from the analyzer. Refer to
Figure
1.
Note: Only analyzers with the auto calibration option have a calibration solution bottle.
b. Discard the solutions in the bottles into an applicable drain.
c. Flush the bottles with deionized water.
d. Fill the calibration solution bottle with new 10-ppm Sodium Chloride (NaCl) solution (500 mL).
The calibration solution bottle has a label with a yellow stripe.
Note: To prepare 10-ppm Sodium Chloride (NaCl) solution, refer to the operations manual.
e. Fill the reactivation solution bottle with new 0.5M Sodium Nitrate (NaNO
The calibration solution bottle has a label with a red stripe.
Note: To prepare 0.5M Sodium Nitrate (NaNO
f. Install the bottles in the analyzer. Make sure that the tubing with a red "REACT" label is put in
the bottle that has a label with a red stripe.
Make sure that the tubing with a yellow "CAL" label is put in the bottle that has a label with a
yellow stripe.
4. Enter the bottle levels as follows:
a. Push menu, then select REAGENTS/STANDARDS.
b. Select SET REACT LEVEL, then enter 100%.
c. Select SET STANDARD LEVEL, then enter 100%.
5. Select PRIME REAGENTS.
6. When the prime reagents step is complete, push menu, then select START ANALYZER.
Figure 1 Bottle removal
) solution, refer to the operations manual.
3
) solution (500 mL).
3
English 5

Publicité

Table des Matières
loading

Table des Matières