Hach NA5600 sc Na+ Maintenance Et Dépannage page 17

Masquer les pouces Voir aussi pour NA5600 sc Na+:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 29
Figure 6 Set the power switch to off
3.9.8 Discard the deionized water in the bottles
1. Remove the reactivation solution and calibration solution bottles from the analyzer. Pull the bottle
caps down so they hang below the analyzer.
Do not let the bottle cap or internal tubing touch the analyzer to prevent contamination.
2. Discard the deionized water in the bottles.
3. Let the bottles fully air dry.
4. Install the bottles in the analyzer.
5. Close the analyzer door.
Section 4 Troubleshooting
Problem
The status bar does
not show on the
measurement
screen.
A channel does not
show on the
measurement screen
when the UP or
DOWN arrow key is
pushed.
Possible cause
The channel shown on
the measurement
screen is not the
channel being
measured.
The channel is not
selected in the display
settings.
Solution
Push the UP or DOWN arrow until the channel
being measured shows on the measurement
screen.
Push menu, then select SETUP SYSTEM >
DISPLAY SETUP > ADJUST ORDER > SEE
CURRENT ORDER.
If the channel is not shown, push back, then
select ADD MEASUREMENTS. Push the LEFT
arrow to select or unselect a checkbox.
English 17

Publicité

Table des Matières
loading

Table des Matières