Technical Specifications; Technical Characteristics - koban KMIR-01 Mode D'emploi

Multimètre thermographique trms
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

KMIR-01

7. Technical specifications

7.1. Technical characteristics

• Thermal imager
Field of view (FOV) / Minimum focus
21° x 21° / 0,5 m
distance
Spatial resolution (IFOV)
4,53 mrad
IR resolution
80 × 80 pixels
Thermal sensitivity/NETD
< 0.1°C @ +30°C (+86°F) / 100 mK
Image frequency
50Hz
Focus mode
Focus free
Focal length
7.5mm
Focal Plane Array (FPA)/ Spectral range Uncooled microbolometer / 8–14 μm
Object temperature range
–20°C to +260°C (–4°F to +500°F)
±3°C (±5.4°F) or ±3% of reading
Accuracy
(Environment temperature 10°C-35°C,
object temperature >0°C. )
Accuracy calculated as [%reading + (num. digits*resolution)] at 18°C ÷ 28°C <75%HR
• DC Voltage
Range
Resolution
Accuracy
±(0.8%reading
400 mV
0,1 mV
+ 8digits)
4 V
0,001 V
±(0.5%reading
+ 5digits)
40 V
0,01 V
400 V
0,1 V
±(0.8%reading
+ 5digits)
600 V
1 V
30
EN
Input
Protection against
impedance
overcharge
> 10 MΩ
600VDC/ACrms
EN
• AC TRMS Voltage
Range Resolution
Accuracy(*)
(50Hz÷60Hz)
(61Hz÷1kHz)
4 V
0,001 V
40 V
0,01 V
±(1.0%reading +
±(2.5%reading +
5digits)
5dgt)
400 V
0,1 V
600 V
1 V
(*) Accuracy specified from 10% to 100% of the measuring range, sine wave.
Input impedance: >9MΩ;
Accuracy PEAK function: ±10%rdg, PEAK response time: 1ms
• AC+ DC TRMS Voltage
Range
Resolution Accuracy
(50 Hz - 1 kHz)
4 V
0,001 V
40 V
0,01 V
±(2.5%reading
400 V
0,1 V
+20dgt)
600 V
1 V
• DC Current
Range
Resolution Accuracy
400 uA
0,1 uA
4000 uA 1 uA
±(1.5%reading + 5digits)
40 mA
0,01 mA
400 mA 0,1 mA
±(1.5%reading + 8digits)
10 A
0,01 A
±(2.0%reading + 8digits)
KMIR-01
Protection against
overcharge
600VDC/ACrms
Input
Protection against
impedance
overcharge
> 10 MΩ
600VDC/ACrms
Protection against
overcharge
fuse 500mA/600V
fuse 10A/600V
31

Publicité

Table des Matières
loading

Table des Matières