7. DISINFECTION, CLEANING AND STERILIZATION:
First clean and decontaminate the instruments and their supports. Then decontaminate and sterilize them out of
their environment (in bags) with the autoclave at 134°C and 3 bar during 18 minutes.
7.1 General recommendation
-
Single use marked instruments are not approved for re-use.
-
The user is responsible for the sterility of the product for the fi rst cycle and each further usage as well as for
the usage of damaged or dirty instruments.
-
For your own safety, please wear personal protective equipments (gloves, glasses).
-
Use only disinfecting solution which is approved for their effi cacy (DGHM-listing, CE marking, FDAapproval)
-
Tungsten Carbide burs, Plastic Support, Hand instruments and NiTi instruments are degraded by Peroxide
hydrogen (H2O2) solution.
-
NiTi Instruments are degraded if immerged more than 5 minutes in a solution of NaOCl at more than 5 %.
-
Instruments made of aluminium are degraded in presence of caustic soda solutions with mercury salt.
-
Do not use acid (pH < 6) or alkaline (pH > 8) solutions.
-
After 5 cycles, degradation of marking may appear.