Chlorine, With Powder Pack (Cl 2) - YSI Chlorine 900 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Methods
CL 2
Tablet
Powder
0.0.0
Tablet
Powder
Zero
Test
CL 2
Tablet
Powder
RESULT
Tablet
Powder
!
Zero
Test
CL 2
Tablet
Powder
RESULT
Tablet
Powder
12
Chlorine with powder pack: 0.02 – 2.0 mg/l
a) Free Chlorine
Fill a clean vial (24 mm Ø) with 10 ml of the water sample and perform zero
calibration (see "Operation").
Remove the vial from the sample chamber.
Add the contents of one Chlorine Free-DpD/ F10 powder pack straight from the
foil into the water sample.
Close the vial tightly with the cap and invert several times to mix the contents (20
sec.).
Place the vial in the sample chamber making sure that the marks are aligned.
Press the [ZERO/TEST] key.
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l free Chlorine.
b) total Chlorine
Remove the vial from the sample chamber.
Rinse the vial and the cap several times and then fill the vial with
10 ml of water sample (Note 2 and 3).
Add the contents of one Chlorine total-DpD/ F10 powder pack straight from the
foil into the water sample.
Close the vial tightly with the cap and invert several times to mix the contents (20
sec.).
Place the vial in the sample chamber making sure that the marks are aligned.
Wait for a reaction period of 3 minutes.
(Countdown can be activated, see page 7)
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l total Chlorine.
c) Combined Chlorine
combined Chlorine = total Chlorine – free Chlorine
Tolerances:
0 – 1 mg/l: ± 0.05 mg/l
> 1 – 2 mg/l: ± 0.10 mg/l

Publicité

Table des Matières
loading

Table des Matières