The CA 6501 is an electrical measuring instrument intended for the measurement
of low resistance values (0 to 100Ω at 5mA
1mA
), and for insulation measurements (0.5 to 200MΩ at 500V
DC
ohmmeter function is protected against the accidental presence of a voltage by a
high-breaking-capacity fuse. A voltage-present indicator serves to make sure that
there is no AC voltage in the circuit to be tested.
This instrument is ready for use at all times because it requires no operating
maintenance. A magneto generator provides the various measurement and
operating voltages.
2 DESCRIPTION OF THE VARIOUS PARTS,
Physical guard used
to support the left hand
when operating the
crank.
Red indicator
lit: multiply the reading
by 10 to obtain the result
(on MΩ, kΩ, Ω scales)
Range switch;
set back to
MΩ/V∼ after using
C.A 6501

1 PRESENTATION

CONTROLS, AND FUNCTIONS
Measurement terminals
Black
), of resistances (0 to 500kΩ at
DC
Fuse
Red
Screw for adjustment
of the mechanical zero
of the galvanometer
English
). The
DC
Orange indicator lit: indicates
that magneto speed of rotation
is OK, i.e. that there is a voltage
of up to 500V
(MΩ range) or
DC
a current of 1mA
(kΩ range)
DC
or a current of 5mA
DC
Retractable crank
17
(Ω range).