Display  
"Temperature
Out of Range"  
"Temp Out of Range
Invalid Test"  
"Test Cancelled"  
"Thermistor Failure"  
"Time Error"  
">XX units"  
"<XX units"  
65000159   R ev.   0 07   ( 06/15)                                                                                                                                                                                                                                                                                                                                                                                                                               5 9  
Troubleshooting the FastPack
Cause  
•   T emperature   t oo   l ow   m ay   o ccur   i f  
door   i s   l eft   o pen   f or   l ong   d uration.  
•   T emperature   t oo   h igh   m ay   o ccur   i f  
room   t emperature   i s   t oo   w arm   ( >32°C  
or   9 0°F).  
Temperature   h as   e xceeded   c orrect  
range   d uring   a   t est.  
Cancel   w as   p ressed   d uring   a   t est.  
Thermistor   h as   f ailed.  
A   t ime   t hat   i s   n ot   p ossible   h as   b een  
entered.  
®
FastPack
  I P   T est   r esult   i s   h igher   t han  
the   m aximum   r eportable   r esult.  
®
FastPack
  I P   T est   r esult   i s   l ower   t han   t he  
minimum   r eportable   r esult.  
IP System
®
Action  
•   C lose   d oor.   R echeck   a fter   5   m inutes.  
  If   p roblem   p ersists,   c ontact   S ystem  
Support.  
•   M ove   a nalyzer   t o   a   c ooler   l ocation   o r  
wait   a pproximately   1 5   m inutes  
between   r uns.  
  If   p roblem   p ersists,   c ontact   S ystem  
Support.  
Discard   p ack.   R etry   t est.   I f   p roblem  
persists,   c ontact   S ystem   S upport.  
Cancelling   a   t est   w ill   r equire   t he   u se   o f   a  
®
new   F astPack
  I P   a nd   n ew   s ample.  
Contact   S ystem   S upport.  
Re-­‐enter   c orrect   t ime.  
Report   r esult   a s   > XX   u nits.  
If   d esired   s end   s ample   t o   r eference   l ab  
for   c onfirmation.  
Report   r esult   a s   < XX   u nits.  