Display
"Temperature
Out of Range"
"Temp Out of Range
Invalid Test"
"Test Cancelled"
"Thermistor Failure"
"Time Error"
">XX units"
"<XX units"
65000159
R ev.
0 07
( 06/15)
5 9
Troubleshooting the FastPack
Cause
•
T emperature
t oo
l ow
m ay
o ccur
i f
door
i s
l eft
o pen
f or
l ong
d uration.
•
T emperature
t oo
h igh
m ay
o ccur
i f
room
t emperature
i s
t oo
w arm
( >32°C
or
9 0°F).
Temperature
h as
e xceeded
c orrect
range
d uring
a
t est.
Cancel
w as
p ressed
d uring
a
t est.
Thermistor
h as
f ailed.
A
t ime
t hat
i s
n ot
p ossible
h as
b een
entered.
®
FastPack
I P
T est
r esult
i s
h igher
t han
the
m aximum
r eportable
r esult.
®
FastPack
I P
T est
r esult
i s
l ower
t han
t he
minimum
r eportable
r esult.
IP System
®
Action
•
C lose
d oor.
R echeck
a fter
5
m inutes.
If
p roblem
p ersists,
c ontact
S ystem
Support.
•
M ove
a nalyzer
t o
a
c ooler
l ocation
o r
wait
a pproximately
1 5
m inutes
between
r uns.
If
p roblem
p ersists,
c ontact
S ystem
Support.
Discard
p ack.
R etry
t est.
I f
p roblem
persists,
c ontact
S ystem
S upport.
Cancelling
a
t est
w ill
r equire
t he
u se
o f
a
®
new
F astPack
I P
a nd
n ew
s ample.
Contact
S ystem
S upport.
Re-‐enter
c orrect
t ime.
Report
r esult
a s
> XX
u nits.
If
d esired
s end
s ample
t o
r eference
l ab
for
c onfirmation.
Report
r esult
a s
< XX
u nits.