Troubleshooting the FastPack
Display  
"Unable to Read
Barcode"  
"Invalid Label"
"Unknown Test"  
"Restricted test type
due to temp"  
"Missing Pack"  
"Motor Failure"  
"PMT Background
High _ _ _"
"PMT Failure"  
"Shutter Failure"  
"Pressure Failure"  
"Remove FastPack"  
58                                                                                                                                                                                                                                                                                                                                                                                                                             6 5000159   R ev.   0 07   ( 06/15)  
IP System
®
Cause  
•   B ar   c ode   l abel   i s   i llegible.  
•   R eader   n on-­‐functional.  
•   W rong   t est-­‐type   p ack.  
•   C orrect   s oftware   v ersion   n ot   i nstalled.  
Operator   a ttempted   t o   r un   a   F astPack
test   t hat   i s   i ncompatible   w ith   t he  
temperature   o f   t he   a nalyzer.  
There   i s   n o   p ack   i n   t he   a nalyzer.  
Motor   h as   f ailed.  
PMT   b ackground   r eading   i s   h igh   d uring  
self-­‐diagnostic   o r   a t   b eginning   o f   a   t est.  
This   m ay   r esult   f rom   c ontamination,  
leakage   o f   a mbient   l ight   i n   t he   a nalyzer,  
or   a ttempting   t o   u se   a   p reviously   u sed  
FastPack®   I P.  
PMT   F ailure;   L ight   L eak.  
Shutter   F ailure  
System   h as   a   s evere   p ressure   l eak.   A ir  
pump   n ot   f unctioning.  
®
A   F astPack
  I P   i s   i n   t he   a nalyzer   w hen  
attempting   t o   r un   d iagnostics.  
Action  
•   C heck   t o   s ee   i f   b ar   c ode   i s   d efaced.  
•   R eposition   b arcode   l abel   a pprox.   ⅛ -­‐
inch   l ower   a nd   r e-­‐try.  
• If   P lexiglas   b arcode   w indow   i s   p resent,  
use   a   s tandard   f lat   h ead   s crewdriver   t o  
remove   t he   w indow   a nd   r e-­‐try.  
®
•   V erify   t hat   F astPack
  I P   a nd   C alibrator  
card   a re   o f   t he   s ame   t est-­‐type.  
•   C ontact   S ystem   S upport  
®
 
Attempt   t o   r un   t he   F astPack
analyzer   t hat   i s   t emperature   c ompatible.  
®
Insert   F astPack
  I P.  
Run   d iagnostic.   I f   i t   f ails,   c ontact   S ystem  
Support.  
Ensure   t hat   t he   u sed   F astPack
still   i n   t he   I nstrument.   E nsure   t hat   t he  
instrument   i s   n ot   i n   d irect   s unlight.  
Follow   c leaning   p rocedure.   R un   s elf-­‐
diagnostic.   C ontact   S ystem   S upport   i f  
message   p ersists.  
Ensure   t hat   t he   a nalyzer   i s   n ot   i n   d irect  
sunlight.   C ontact   S ystem   S upport.  
Contact   S ystem   S upport.  
Contact   S ystem   S upport.  
®
Remove   F astPack
  I P.   U pon   c losing   d oor,  
diagnostics   w ill   a utomatically   b egin.  
®