Running Patient Samples
When to Run a FastPack
The FastPack
IP System analyzer is ready to run a FastPack
®
reached operating temperature and pressure, and has been calibrated. If the analyzer is not at operating
temperature, the analyzer will not operate and the BLUE button will not function. Leave the analyzer
turned on 24 hours a day so that it will always be at operating temperature.
The analyzer performs a comprehensive Self Diagnostic test every midnight to ensure the system is
functioning properly. When the Self Diagnostic test is running, the analyzer door will be locked shut.
What You Need
1. FastPack
IP System
®
2. Microcentrifuge (required for patient samples
only)
3. FastPack IP (for the specific assay to be tested)
4. Pipette (available from Qualigen)
5. Pipette Tips (available from Qualigen)
6. Patient Sample (serum or plasma depending on
the assay)
7. Medical Gloves - used to avoid contact with
biohazard materials such as human blood, etc.
FOR FASTPACK
8. Pretreatment Buffer