Cleaning the FastPack
IP System
®
Note: Always wear protective gloves when cleaning or
disinfecting the analyzer.
The FastPack
IP System analyzer is completely self-
®
enclosed, and only requires periodic cleaning to remove
excess sample. Use a damp cloth to wipe down the entire
exterior and interior compartment of the analyzer. Do
not use solvents to wipe down the exterior of the
analyzer.
If a FastPack
IP has leaked inside the FastPack
IP
®
®
analyzer, perform the following cleaning procedure:
 
FastPack
IP System Cleaning Procedure
®
Action  
Illustration  
Wipe   t he   i nside   o f   t he   d oor   p anel   w ith   a   s oft   c loth   s oaked   w ith   a   m ild   d isinfectant   o r  
any   h ospital   g rade   d isinfectant   t o   r emove   a ny   p otential   c ontaminant   t hat   m ay  
impact   p erformance   o f   t he   a nalyzer.  
Note:   D o   n ot   p ress   t oo   h ard   a s   t he   s ilicone   m embrane   c an   b e   t orn   w ith   e xcessive  
force.  
 
Wipe   t he   m embrane   s ide   o f   t he   a nalyzer   c hamber   w ith   a   m ild   d isinfectant   t o  
eliminate   a ny   p otential   a ctive   c hemiluminescence   c hemicals   t hat   m ay   i mpact  
performance.  
 
65000159   R ev.   0 07   ( 06/15)                                                                                                                                                                                                                                                                                                                                                                                                                               5 5  