RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND
MOBILE RF COMMUNICATIONS EQUIPMENT AND COMPRESSOR NEBU-
LIZER AGU N3
Compressor nebulizer AGU N3 is intended for use in an electromagnetic environ-
ment in which radiated RF disturbances are controlled. The customer or the user
of compressor nebulizer AGU N3 can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communi-
cations equipment (transmitters) and compressor nebulizer AGU N3 as recom-
mended below, according to the maximum output power of the communications
equipment. For transmitters rated at a maximum output power not listed above,
the recommended separation distance d in meters (m) can be determined using
the equation applicable to the frequency of the transmitter, where P is the maxi-
mum output power rating of the transmitter in watts (W) according to the trans-
mitter manufacturer.
Rated maximum
output power of
transmitter (W)
0.01
0.1
1
10
100
NOTE 1: at 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: these guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
WARNING
This symbol on the product means that this product is an electronic device
which must be disposed of in accordance with Directive 2012/19/EU at the
local waste disposal utility.
EN
Separation distance according to frequency of transmitter (m)
150 kHz – 80 MHz
d = 1.2 √P
0.12
0.38
1.2
3.8
12
80 MHz – 800 MHz
d = 1.2 √P
0.12
0.38
1.2
3.8
12
800 MHz – 2.5 GHz
d = 2.3 √P
0.23
0.73
2.3
7.3
23
61