TECHNICAL DETAILS

Measurements of the particle size using the cascade impactor for AGU N3 accord-
ing to BS EN 13544-2009 «Respiratory equipment – Part 1: Nebulizing systems
and their components», Annex CC.3, based on the measurements of the particle
size with the cascade impactor are as follows:
Size of nebulized particles
Capacity of medicine cup
Distribution of particles
40.00
35.00
30.00
25.00
20.00
15.00
10.00
5.00
0.00
0.00 0.52 0.93 1.55 3.50 6.00 9.80 14.80 21.30
Particle size (µm)
EN
MMAD ≤ 2.9 µm
(MMAD – Mass Median Aerodynamic
Diameter)
Max. 6 ml