IEC 60601
Immunity test
Test level
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
Electrical fast
± 2 kV for power
transient/burst
supply lines
IEC 61000-4-4
± 1 kV for input/
output lines
Surge
± 1 kV line(s)
IEC 61000-4-5
to line(s)
± 2 kV line(s)
to earth
Voltage
<5% U
(>95% dip in U
dips, short
for 0.5 cycle
interruptions
and voltage
40% U
variations on
(60% dip in U
power supply
for 5 cycles
input lines
70% U
IEC 61000-4-11
(30% dip in U
for 25 cycles
<5% U
(>95% dip in U
for 5 s
Power frequency
3 A/m
magnetic fields
(50/60 Hz)
IEC 61000-4-8
NOTE: U
– is the AC mains voltage prior to application of the test level.
T
EN
Compliance
level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
Not applied
± 1 kV line(s)
to line(s)
Not applied
<5% U
T
T
)
(>95% dip in U
T
for 0.5 cycle
40% U
T
T
)
(60% dip in U
T
for 5 cycles
70% U
T
T
)
(30% dip in U
T
for 25 cycles
<5% U
T
T
)
(>95% dip in U
T
for 5 s
3 A/m
Electromagnetic
environment – guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the rela-
tive humidity should be at
least 30 %
Mains power quality should
correspond to the standards
of the public power supply
networks
Mains power quality should
correspond to the standards
of the public power supply
networks
Mains power quality should
)
correspond to the standards
T
of the public power supply
networks.
If the user of the device
)
T
requires continued operation
during power mains interrup-
tions, it is recommended that
)
AGU N3 to be powered from
T
an uninterruptible power
supply or a battery
)
T
Power frequency magnetic
fields of AGU N3 should be
at levels characteristic of a
typical public use systems
59