► Use caution when working with the product.
► If the product is damaged, check it for proper function and readi­
ness for use.
► In case of changes in or loss of functionality, do not continue
using the product (see "Signs of changes in or loss of functional­
ity during use" in this section).
► Take any necessary measures (e.g. repair, replacement, inspec­
tion by the manufacturer's customer service, etc.).
Signs of changes in or loss of functionality during use
Decreased forefoot resistance or changes in roll-over behaviour are
noticeable indications of loss of functionality.
4 Scope of delivery
Quantity
1
1
Spare parts/accessories (not included in the scope of deliv­
Designation
Connection cap
5 Preparing the product for use
CAUTION
Incorrect alignment or assembly
Risk of injury due to damaged prosthetic components
► Observe the alignment and assembly instructions.
NOTICE
Grinding the prosthetic foot or footshell
Premature wear resulting from damage to the product
Designation

Instructions for use

Prosthetic foot
ery)
Reference num­
ber
2C10*, 2C11*
► Do not grind the prosthetic foot or footshell.
5.1 Alignment
5.1.1 Bench Alignment
TT bench alignment
Bench alignment process
Required materials: 662M4 goniometer, 743S12 heel height
measuring apparatus, 743A80 50:50 gauge, alignment apparatus
(e.g. 743L200 L.A.S.A.R. Assembly or 743A200 PROS.A.
Assembly)
Perform the assembly and alignment of the prosthetic components
in the alignment apparatus according to the following specifica­
tions:
Sagittal plane
Heel height: Effective heel height (shoe heel height - sole
   
thickness of forefoot) + 5 mm
Exterior foot rotation: approx. 5°
   
   
Anterior placement, middle of the prosthetic foot to the align­
ment reference line: 30 mm
   
Connect the prosthetic foot and prosthetic socket using the
chosen adapters. The instructions for use of the adapters
must be observed.
   
Determine the centre of the prosthetic socket with the 50:50
gauge. Align the prosthetic socket centrally to the alignment
reference line.
Socket flexion: Individual residual limb flexion + 5°
Frontal plane
   
Alignment reference line of prosthetic foot: between big toe
and second toe
Alignment reference line of prosthetic socket: along the lat­
eral patella edge
   
Observe the abduction or adduction position.