Storage And Transport - lopital Amfora Mode D'emploi

Brancard de douche / table d'habillage
Masquer les pouces Voir aussi pour Amfora:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Maintenance schedule
After every use
Weekly
Annually

6. Storage and transport

During transport and storage, the ambient conditions must remain moderate. In other
words, the temperature should remain between -40°C and +55°C (-40°F to 130°F) and
humidity levels should be kept as low as possible. Vibration during transport must be kept
to a minimum in order to avoid damaging the hinged mechanisms on the stretcher.
44 | Instructions for use | Shower stretcher / Dressing table
• The shower stretcher / dressing table can be cleaned with
water and regular household cleaning products.
• Clean and disinfect mattress / harness.
• Clean frame.
• Check wheels for hairs, dust and soap residues and remove
if necessary.
• Check that the shower stretcher / dressing table is still
working correctly.
• According to the guidelines for medical devices, a shower
stretcher / dressing table must be maintained in line with
the manufacturer's instructions. This process will preferably
be carried out by Lopital or another authorised body. Lopital
offers the option of having the shower stretcher / dressing
table thoroughly inspected and serviced once a year via
Periodic Preventative Maintenance.
For more information about the maintenance contract,
please contact Lopital (service@lopital.nl) or speak to
your supplier.