portable and mobile RF communications equipment and the shower stretcher / dressing table
The shower stretcher / dressing table is intended for use in the electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the shower stretcher / dressing table can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the shower
stretcher / dressing table as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watt (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
Brancard de douche / Table d'habillage | Mode d'emploi | 79
Recommended separation distance between
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d =1,2 √P
0,12
0,38
1,2
3,8
12
m
80 MHz to 800 MHz
800 MHz to 2,5 GHz
d =1,2 √P
d =2,4 √P
0,12
0,24
0,38
0,76
1,2
2,4
3,8
7,6
12
24