Meter calibration data
error.
The component of
the device might be
Review the instructions
broken.
and repeat the test with
a new strip and using
the correct technique.
If the problem persists,
contact your local
customer service for
Meter malfunction
assistance.
due to damaged
components.
Incorrect data on the
meter or the code
strip.
You may have
removed the strip
after applying blood,
or there is insufficient
blood volume.
Ambient
temperature is
below the operating
temperature.
Ambient
temperature is
above the operating
temperature.
Appears when the
test strips are expired.
(This only applies
to β-ketone, total
cholesterol and uric
acid test strips)
EN-20
Review the instructions
and repeat the test with
a new test strip. Please
make sure you use
the correct technique
and dose with enough
blood.
The operating
temperature range is
from 8°C to 45°C (46.4°F
to 113°F). Repeat the
test after the device and
test strip have reached
the above-mentioned
operating temperature.
Ensure the date and
time settings of the
meter are correct and
up to date. Insert the
code strip again. If the
problem persists, please
check the expiry date of
the test strip.
Discard the expired test
strips. Use a new test
strip that is in date and
repeat the test.