Continuous Display Of Measured; Result (Hold) - Seca 704 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 71
Continuous display of
measured

result (HOLD)

4. Hold down the arrow key (hold/2 in 1) until NET
appears in the display.
The weight is stored.
0. 0
appears in the display.
ATTENTION!
Incorrect measurement due to change in initial
weight
If the child is weighed with a different initial
weight, the child's weight will not be calculated
correctly.
– Make sure that the child is always weighed
with the adult used to determine the initial
weight.
– Check that the weight of this adult does not
change, e.g. by taking off a garment.
5. Ask the adult to step onto the scales with the child.
The child's weight is displayed.
The
symbol, the
"HOLD" and "NET" are displayed.
6. Ask the adult to step off the scales with the child.
7. To disable the 2 in 1 function, press the arrow key
(hold/2 in 1) until the
and the messages "HOLD" and "NET" are no longer
displayed or switch the scales off.
If you activate the HOLD function, the weighing value
will continue to be displayed after the load has been
removed. This allows you to attend to the patient
before noting down the weight.
1. Check that there is no load on the scales.
2. Ask the patient to step onto the scales.
3. Briefly press the arrow key (hold/2 in 1).
The display will flash until a stable weight is meas-
ured. The weighing value will then be displayed
continuously. The
tion) and the message "HOLD" will be displayed.
4. To disable the HOLD function, briefly press the ar-
row key (hold/2 in 1).
The
symbol and the message "HOLD" will no
longer be displayed.
symbol and the messages
symbol, the
symbol and
symbol (non-calibratable func-
Operation • 49
English

Publicité

Table des Matières
loading

Table des Matières