SYMPTOM
Unit does not operate
No Display
Display shows  
Display shows  
Alarm light on
Temperature does not rise
when expected
SYMPTOM
Temperature continues to
rise when not expected
Stirred motor does not ro-
tate
17090 Version 15 - June 2009 83S
Page 19
POSSIBLE CAUSE
Unit not switched on
Unit not plugged into power
supply
Fuse blown in the unit or for UK
units, in the plug
Electrical power supply failure
Probe open circuit
Probe short circuit
Overtemperature cut-out has
operated (GD120)
Low liquid level, float switch
has operated
Set temperature is lower than
liquid temperature
Set temperature is too close to
ambient temperature
Temperature control circuit fault
POSSIBLE CAUSE
Set temperature is higher than
liquid temperature
Temperature control circuit fault
Stirrer or propeller obstructed
Pump obstructed
If stirrer motor shaft rotates
freely then motor thermal fuse
may have blown
www.grant.co.uk

4. Fault diagnosis

ACTION REQUIRED
Switch on
Plug in, switch on
Check and replace
Check that other electrical
appliances on the same
circuit are working
Have unit checked by a
competent person
Have unit checked by a
competent person
Reset cut-out
Top up liquid
Check set temperature
Raise the set temperature or fit
accessory cooling
Have unit checked by a
competent person
ACTION REQUIRED
Check set temperature
Have unit checked by a
competent person
Clear obstruction
Clear obstruction
Have unit checked by a
competent person
GD100, GD120