Gd100 And Gd120 Preset Temperatures - Grant GD100 Instructions D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

3.3.2
GD100.
Switch the power on Press the S button and rotate the "Navigator" knob to set
the required operating temperature. Press S button again to fix this value.
3.4
GD100 and GD120 (Fig. 2) Preset temperatures
Setting preset temperature values
Press F, Rotate the "Navigator" knob until required pre set adjust number [,  or ]
indicated on the display. (The display will alternate between the „‟ and the preset
is
temperature). Press S Rotate the Navigator knob until desired temperature value is
indicated on the display. Press S. Controller accepts this as current set temperature
value and also stores it as a preset value. Running at preset temperature value Press F
Rotate the set temperature control until required pre set number [,  or ] is indicated
on the display. Press S Press S Controller accepts this as current set temperature.
Preset temperatures can be adjusted over full range of unit (-20° to 120°C), but will only
be accepted when over 100°C if set to [] and below 100°C when set to low
temperature liquid [] When set to water [], preset values greater than 100°C will
revert to 100°C when S is pressed to leave preset menu.
17090 Version 15 - June 2009 83S
Page 13
www.grant.co.uk
GD100, GD120
Operating Manual

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Gd120

Table des Matières