Completion Of Patient Placement; Home Display; Instaflate ™ Inflation Function (If) - Arjohuntleigh First Step Select Manuel De L'utilisateur

Système de remplacement de matelas
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Completion of Patient Placement

Home Display

1.
Verify patient comfort.
2.
Use Home Display for up to 25% uniform pressure adjustment.
• press left arrow button under soft to decrease uniform pressure.
• press right arrow button under fIrm to increase uniform pressure.
3.
If the decision has been made to use side rails, raise and lock side rails.
serious injury or death can result from the use (potential entrapment) or non-use (potential patient falls) of side rails
or other restraints.
Side Rails / Patient Restraints - Whether and how to use side rails or restraints is a decision that should be based on
each patient's needs and should be made by the patient and the patient's family, physician and caregivers, with facility
protocols in mind. Caregivers should assess risks and benefits of side rail / restraint use (including entrapment and
patient falls from bed) in conjunction with individual patient needs, and should discuss use or non-use with patient
and / or family. Consider not only the clinical and other needs of the patient but also the risks of fatal or serious injury
from falling out of bed and from patient entrapment in or around the side rails, restraints or other accessories. In the
Us, for a description of entrapment hazards, vulnerable patient profile and guidance to further reduce entrapment
risks, refer to fDa's Hospital Bed system Dimensional and assessment Guidance to reduce entrapment. outside the
Us, consult the local Competent authority or Government agency for medical Device safety for specific local guidance.
Consult a caregiver and carefully consider the use of bolsters, positioning aids or floor pads, especially with confused,
restless or agitated patients. It is recommended that side rails (if used) be locked in the full upright position when
the patient is unattended. make sure a capable patient knows how to get out of bed safely (and, if necessary, how
to release the side rails) in case of fire or other emergency. Monitor patients frequently to guard against patient
entrapment.
When selecting a standard mattress, ensure the distance between top of side rails (if used) and top of mattress
(without compression) is at least 8.66 in (220 mm) to help prevent inadvertent bed exit or falls. Consider individual
patient size, position (relative to the top of the side rail) and patient condition in assessing fall risk.
i
adjust bed height to lowest level.
to minimize risks of injury from falls, the bed frame should always be in the lowest possible position when the patient
is unattended.
4.
ensure all brakes are locked.
5.
activate warmer and adjust temperature if desired.
InstaFlate
Inflation Function (IF)
Use Instaflate to assist patient transfer and bathing by increasing pressure throughout surface. Instaflate will automatically
deactivate after 15 minutes. screen will flash every second in the last 15 seconds before deactivation.
1.
Press the INSTAFLATE ON/OFF button shown at right.
Display shown at right will appear and flash until Instaflate is deactivated.
2.
Press the INSTAFLATE ON/OFF button to deactivate InstaFlate.
Display shown at right will appear temporarily and then return to the Home Display.
pressure and warmer settings will be restored when Instaflate is deactivated.
INSTAFLATE
INSTAFLATE OFF
13

Publicité

Table des Matières
loading

Table des Matières