How To Enter The Rescue Pin Code - Knick Portavo 908 MULTI Instructions

Masquer les pouces Voir aussi pour Portavo 908 MULTI:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

48
pH
ORP

How to Enter the Rescue PIN Code

1) In measuring mode, press the Menu softkey.
2) Select "Configuration" using the cursor keys.
3) Press the  and  keys simultaneously.
4) Set User to "ADMIN".
5) Select "PIN code" and enter the rescue PIN code.
Confirm by pressing enter.
6) Press the Continue softkey.
Sensor Verification
To make sure that only selected sensors can be operated on the meter, you can eval-
uate the following data stored in the sensor:
- Model (sensor model)
- TAG (e.g., point of measurement)
- Group (e.g., facility)
With Option 001 enabled, the "Configuration" menu is extended as follows:
- Sensor verification
Check model
Check TAG
Check group
You can select the following options:
Off
No verification.
Info
When a wrong sensor is connected, an error message will be
displayed.
Nevertheless, you can continue working with the sensor.
Reject
Here you specify values with which the sensor will be rejected.
Oxy Cond
Off | Info | Reject
Off | Info | Reject
Off | Info | Reject
Options