Daily/After each use:
1. Disassemble the mask according to the disassembly instructions.
2. If required, wipe your cushion with a CPAP wipe or equivalent and
then allow to dry. Important – do not get cushion wet.
3. Rinse the frame and elbow under running water. Clean with a soft
brush until dirt is removed.
4. Soak the components in warm water with a mild liquid detergent for
up to ten minutes.
5. Shake the components in the water.
6. Brush the moving parts of the elbow and around the vent holes.
7. Brush the areas of the frame where the arms connect, and inside and
outside the frame where the elbow connects.
8. Rinse the components under running water.
9. Leave the components to air dry out of direct sunlight. Make sure to
squeeze the arms of the frame to ensure that excess water is
removed.
Weekly:
1. Disassemble the mask. The magnets can remain attached to the
headgear during cleaning.
2. Handwash the headgear in warm water with mild liquid detergent.
3. Rinse the headgear under running water. Inspect to ensure the
headgear is clean and detergent free. Wash and rinse again, if
necessary.
4. Squeeze the headgear to remove excess water.
5. Leave the headgear to air dry out of direct sunlight.
Monthly:
Replace the cushion by following relevant steps of the disassembly /
reassembly instructions.
Reprocessing the mask between patients
Only Sleep Lab Mask (SLM) variants of the mask are intended for
multi-patient re-use. When using between patients, these masks must be
reprocessed according to cleaning and disinfection instructions available
on ResMed.com/downloads/masks.